Design Based On Fundamental NIST Developed Principles
A system capable of continuous, high-accuracy humidity generation over a wide range of humidity, temperature, and flow rates.
The CGS-240’s specifications are outlined, below
| Generating Range: | |||
|---|---|---|---|
| Relative Humidity | 10 … 95 %RH (5 %RH optional) | ||
| Dew/Frost Point | -25 … 70 °C DP, -23 … 0 °C FP | ||
| Chamber Temperature | -0 … 72 °C (-10 & +85 °C extended ranges optional) | ||
| Chamber Pressure | Ambient | ||
| Flow Rate | 5 … 20 SLPM (5 … 50 SLPM optional | ||
| Accuracy: | (0 to 70°C, 10 to 40 l/min) | ||
| Relative Humidity | The larger of ±0.1 %RH or ±0.5% of indicated RH value | ||
| Dew/Frost Point | ±0.10 °C | ||
| Temperature (TS, TC) | ±0.030 °C | ||
| Pressure | ±0.02% of full scale | ||
| Chamber Temp Uniformity | <0.035 °C | ||
| Flow Rate | ±1 SLPM | ||
| Integrated Sensors: | Cal Range: | Cal Uncertainty: | |
| Low Range Pressure (PL) | Ambient … 25psia | ±0.005 psia | |
| High Range Pressure (PH) | Ambient … 150psia | ±0.030 psia | |
| Saturator Temperature (TS) | -15 … 72 °C | ±0.03 °C | |
| Chamber Temperature (TC) | 0 … 72 °C | ±0.03 °C | |
| Chamber Liquid Temp (TCL) | 0 … 72 °C | ±0.10 °C | |
| Presaturator Temp (TP) | 0 … 75 °C | ±0.10 °C | |
| Flowmeter | 0 … 50 SLPM | ±1 SLPM | |
| Chamber: | |||
| Dimensions | 15”H x 12”W x 12”D (380 x 300 x 300 mm) | ||
| Type | Circulating air-jacket encased within liquid jacket | ||
| Adjustable Shelf | Three shelf positions, one shelf supplied | ||
| Access Ports | Three 50 mm diameter thermally controlled access ports | ||
| Window Size | 14”H x 12”W x 12”D (370 x 300 x 300 mm) | ||
| Window thermal control | Liquid jacketed, multi-pane, thermally controlled | ||
| Chamber light | Integrated light with on/off touchscreen control | ||
| Integrated Control Interface: | |||
| Operation Software | RHS Control Software | ||
| Calibration Software | RHS Plateau Software | ||
| Display | 18.5” Multi Touch Monitor with pinch/zoom/drag | ||
| Gas Inlet Pressure: | |||
| Using external pressure source | 120 psig max | ||
| Using supplied air compressor | 60 … 90 psig @ 20 l/min typical | ||
| Environmental Conditions: | |||
| Operating Environment | 15 … 30 °C, 20 … 60 %RH | ||
| Storage Environment | 0 … 50 °C, <95 %RH non-condensing | ||
| Power Requirements: | |||
| Voltage | 208 … 240 VAC, 50/60 Hz, single phase | ||
| Current | 20 Amps | ||
The CGS-240 has specific facility requirements. Please read the entire CGS-240 manual before operating.
Learn more about our CGS-240 Humidity Generator
RH Systems (RHS) warrants the products it manufactures or distributes to be free of defects in material and workmanship under normal use and service when operated within the specified design limitations for a period of 12 months from date of initial shipment. Under this Warranty, RHS will, at its discretion, repair or replace any component that upon examination by RHS or its duly authorized representatives proves to be defective during the warranty period provided the system is returned to the factory for inspection and repair shipping prepaid. Improper or unauthorized maintenance, storage, repair, or alteration of any kind by personnel other than RHS or its duly authorized representatives may void all warranties. Warranty may also be voided for misuse, neglect, accident, corrosion, and improper installation. This Warranty is exclusive and in lieu of any and all other warranties of merchantability, fitness for a particular purpose, or any other warranty, expressed or implied, and all other liabilities and obligations on the part of RHS. RHS will not be liable for any other claims or damages, either direct, indirect, or consequential arising out of the use of its products.
A system capable of continuous, high-accuracy humidity generation over a wide range of humidity, temperature, and flow rates.
The system incorporates a cable management system, sample gas injection into the center of an air circulation fan, local pressure measurement, and a multi-position shelving system. These components are enclosed within high-performance thermal insulation and sealed with a soft silicone profile to ensure a perfectly tight environment.
The CGS-240 may run continuously without the need to shut down (the only requirement being that sufficient water level is maintained in the 20 liter distilled water reservoir). The reservoir may be filled at any time without the need to stop the system, thereby allowing uninterrupted generation of humidity for days, weeks, or months.
RHS Control allows for various instruments to connect via digital communication for continuous monitoring, control, and associated data collection. Data is automatically stored to files during operation for later import to Excel and other post-process analysis and graphing programs.
At RH Systems, we’re constantly pushing ourselves towards advancing our products and services for a wide range of clients, setting a new standard for accuracy and paving the way for future breakthroughs.
Customer satisfaction is the foundation of our reputation, so we never take a one-size-fits-all approach. We work with you to understand your needs and provide customized solutions that work seamlessly right from the jump.
Ready to chat with our Sales Team? Send an email to sales@rhs.com and we’ll get right back to you.
As a NVLAP accredited laboratory, RH Systems is reassessed to ensure that our quality and practices maintain such standards (NVLAP Laboratory Code 600161-0). The National Institute of Standards and Technology (NIST) administers the National Voluntary Laboratory Accreditation Program (NVLAP), which accredits public and private laboratories based on evaluation of their technical qualifications and competence to carry out specific calibrations or tests.
NVLAP provides an unbiased third-party evaluation and recognition of performance. Accreditation signifies that a laboratory has demonstrated compliance with NVLAP management and technical requirements, including quality systems, personnel, accommodation and environment, test and calibration methods, equipment, measurement tractability, sampling, handling of test and calibration items, and test and calibration reports. Learn more here.